Installation guide H-1000-5029-05-BPI 200 interface for the TP200 system
8IntroductionTP200 probe sensorTP200 stylus moduleStylusKinematic couplingSCR200 stylus change rackPI 200 interfaceFigure 1 - TP200 precision touch-tr
9Product description2 Product description2.1 System overviewA small deflection of the stylus tip produces a force which is applied via the stylus mod
10Product descriptionThe timing relationships for an idealised probe signal are indicated in figure 2.A solid state relay (SSR) output is provided whic
11Product descriptionTrigger level 2 and HALTTrigger level 1Probe signalVtDebounce times set by switch 6 and switch 7Delay time set by switch 11 and
12Product description2.4 SYNC and HALT debounceThe SYNC and HALT signals are debounced to prevent spurious triggers occurring as a result of CMM or s
13Product description2.5 Probe damped signal (PDAMP)During high-speed position moves (fast traverse), it is necessary to reduce probe sensitivity to
14Product description2.6 Stylus mass and ambient temperature compensationIn the TP200, probe a reference strain sensor, outside the strain path, adju
15Product description2.8 Front panel indicatorsFigure 6 - PI 200 interface front panelTable 1 - Front panel indicatorsIndicator Colour FunctionPOWER
16Product description2.9 Audible indicatorThe audible indicator has two functions:1. To provide a brief indication that a probe trigger has occurred
17Product description2.11 Configuration switchesCAUTION: Check the PI 200 version number on the rear panel. The switch information in table 2 refers
© 2002 - 2007 Renishaw plc. All rights reserved. Renishaw® is a registered trademark of Renishaw plc.This document may not be copied or reproduced in
18Product descriptionSwitch 1:Inverts the polarity of the HALT output.Switch 2:Selects control of the LEDOFF output to either internal PI 200 control
19Product descriptionFigure 8 - SYNC debounce optionsTrigger levelProbe signalVtSwitch 6 - DownSwitch 7 - DownSYNCSYNCSYNCSYNC20 ms 20 ms adaptive100
20Product descriptionSwitch 10:In operating situations where there is an unusually high level of background vibration, it may be necessary to reduce t
21Product description2.12 Probe polarity switchThe TP200 probe is polarity sensitive and must be connected to the PI 200 such that the centre contact
22Connector pin-outs3 Connector pin-outs3.1 PICS input connectorThe PICS input connector is a 9 pin ‘D’ type socket. The pin numbers are illustrated
23Connector pin-outs987654321Figure 10 - PICS output connector pin numbersTable 5 - PICS output connectorPin numberDescription1 STOP2 PPOFF3 Ground (0
24Connector pin-outs3.3 Solid state relay (SSR) output connectorThe solid state relay output connector is a 7 pin DIN socket. The pin numbers are ill
25Connector pin-outs3.4 Stylus change rack (SCR) output connectorThe SCR200 stylus change rack is connected to the PI 200 via a 6 pin miniature DIN s
26PICS dimensions and cable data4 PI 200 dimensions and cable dataDimensions1/3 rack wide x 2U high 140 mm x 88 mm x 183 mm deepWeight1.25 kgMounting
27Part number summary5 Part number summaryPI 200 interface A-1207-0050 Contents: PI 200 interface unit Installation guide Rubber feet (sel
1PI 200interface for TP200 probe systeminstallation guide
28Part number summary 5.1 ReplacementsTP200 probe system user’s guide H-1000-5014 (English + multilingual CD)PI 200 interface for TP200 probe H-10
29Installation procedure6 Installation procedureWARNING: Ensure the PI 200 is disconnected from the mains supply during installation.WARNING: Take
30Installation procedureFigure 15 - Mounting the PI 200 adjacent to the Renishaw PHC10-2Remove the blanking plugs from the sides of each unit. Separat
31Installation procedureFigure 16 - System interconnection - MH8 headCable 2Cable 1MH8TP200PI 200PICS output CMM controller6.3 System interconnection
320Installation procedureFigure 17 - System interconnection - MIH headCable 2Cable 1MIHTP200PI 200PICS output CMM controllerPAA1 adaptorFigure 18 - Sy
33SerialMADE INUKV.PICS OUTHEADRS23211114HCU15 18OUTPUT10Installation procedureFigure 19 - System interconnection - PH10M headCable 4Cable 3SCR200TP20
34SP600MInstallation procedureFigure 20 - System interconnection - SP600M scanning probe and autochangeCable 4Cable 3AC2PI 200Communications to CMM co
35Installation procedureFigure 21 - System interconnection - TP7M touch-trigger probe and autochangeCable 4Cable 3PI 7-2Communications to CMM controll
36Installation procedure6.4 Using TP200 with the ACR1 autochange systemFigure 22 - Shock absorbing pad fitted to PH10MThe CMM controller must assert t
37Installation procedure6.5 SCR200 stylus change rackThe miniature DIN connector on the base of the SCR200 rack is connected to the ‘SCR200’ connecto
2FCCInformation to user (FCC section 15.105)This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant
38Installation procedure6.6 Interconnection cablesTable 8 - Interconnection cablesCable numberCable ident.Length (m) Part number Notes1PL1 0.26 - 0.7
39Installation procedureTable 8 - Interconnection cables continuedCable numberCable ident.Length (m) Part number Notes10 PL112 1 A-2172-000411 PL97 0.
40PICS terminations7 PICS terminationsThe Renishaw product interconnection system (PICS) has been developed to standardise the format for real-time c
41Summary of changes from earlier versions8 Summary of changes from earlier versionsThe table below summarises the changes made from version V7 - V11
42Maintenance9 Maintenance9.1 PI 200Periodically check the security of mounting screws and electrical connectors.Periodic electrical safety checks s
43Fault finding10 Fault findingTable 10 - The CMM will not register a probe trigger, but the probe operates normally when the stylus is deflected by han
44Fault findingTable 13 - The probe will not arm or the probe does not stay armed when the RESET button is released. The probe LEDs are always ON.PI 20
45Fault findingTable 16 - False (‘air’) triggers occur at traverse speed and the probe LEDs flicker.PI 200 indicators:‘DAMPED’ LED is ON.‘SEATED’ LED op
46Fault findingTable 18 - There is an unexpected loss of accuracy.PI 200 indicators:‘TP200’ LED is ON.‘SEATED’ LED operates normally.Possible causeChec
3Care of equipment Your Renishaw probe and accessories are precision instruments. Please use and maintain the products in accordance with these instru
Renishaw plcNew Mills, Wotton-under-Edge, Gloucestershire, GL12 8JR United Kingdom T +44 (0)1453 524524 F +44 (0)1453 524901 E [email protected]
4SafetySafetyIf this product is not used in its intended manner, any protection provided may be impaired.There are no user serviceable parts inside th
5ContentsContents1 Introduction ...72 P
6Contents7 PICS terminations .......408 Summary of chan
7Introduction1 IntroductionThe TP200 probe system comprises the TP200 probe sensor and stylus module, the PI 200 interface and the optional SCR200 st